iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration ne... — FDA record Z-1751-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 20, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the iSED Automated ESR Analyzer PN: 112-13-002
Reason
Primary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube
Classification
Class II
Recalling firm
Alcor Scientific, Inc.
Quantity
38 units
Agency-reported status
Ongoing
Source record identifier
Z-1751-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot 4/22

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 21, 2022
Recall initiation date
Jul 20, 2022
Center classification date
Sep 15, 2022