Recall archive / FDA device enforcement
Disposable Subdermal Needle Electrode, Pt/lr, 12 x 0.40mm, product number TE/S46/638, 25 needles per package, sterile. — FDA record Z-1737-2022
Reported terminated
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Aug 9, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Disposable Subdermal Needle Electrode, Pt/lr, 12 x 0.40mm, product number TE/S46/638, 25 needles per package, sterile.
- Reason
- Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU.
- Classification
- Class II
- Recalling firm
- Technomed Europe
- Quantity
- 37/25-electrode boxes
- Agency-reported status
- Terminated
- Source record identifier
- Z-1737-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Lot numbers 044220, exp. 1/1/2023; 045077, exp. 3/1/2023; 045443, exp. 4/1/2023; and 049285, exp. 7/1/2024; UDI-DI 08718375861530. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the states of CA, FL, NH, OH, and WA.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Sep 21, 2022
- Recall initiation date
- Aug 9, 2022
- Center classification date
- Sep 12, 2022
- Termination date
- Apr 11, 2024