Artix MT Thrombectomy Device, REF: 32-102 — FDA record Z-1728-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 8, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Artix MT Thrombectomy Device, REF: 32-102
Reason
Thrombectomy device specifies vessel range of 2.5-6 mm on the pouch, carton, and Instructions for Use rather than the correct vessel range of 3-6 mm, which could incrementally increase the radial force on the artery, which has the potential of vascular complications.
Classification
Class II
Recalling firm
Inari Medical - Oak Canyon
Quantity
45
Agency-reported status
Ongoing
Source record identifier
Z-1728-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00810123710584, Lot: 24100082, Expiration: 14 April 2025. IFU: IU-01087 Rev. B

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of FL, IL, WI, KY, TX, CA, OH, GA, AK, MI, PA, WA, CO, OH.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 14, 2025
Recall initiation date
Apr 8, 2025
Center classification date
May 6, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.