iGo¿2 DC Car Adapter to be used with iGo¿2 Portable Oxygen Concentrator Systems. Power cord allows the user to operat... — FDA record Z-1727-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 11, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
iGo¿2 DC Car Adapter to be used with iGo¿2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.
Reason
Potential for DC Power Supply housing to become hot to the touch and deform the plastic housing.
Classification
Class I
Recalling firm
Medical Depot Inc.
Quantity
89,532 units
Agency-reported status
Ongoing
Source record identifier
Z-1727-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityDC Cord Code Info: Catalog No. DV6X-619; UDI 885304020585; Cord Revision: Rev E; Lot No. beginning with 2023-01-01 through 2024-12-31. iGo2 Portable Oxygen Concentrator Systems Code Info: Catalog No. 125D, 125D-XB, 125D-BT, 125D-BT-XB, 125D-ARYA-XB; UDI 885304022237, 885304032250, 885304033257, 885304033240, 885304033516; Lot No. D23125001DS to D23A26121DS, D23706001DA to D23B16056DA, F23B06001DS to F25108143DS, F23313001DH to F25109259DH, F24212001DA to F25217150DA.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Albania, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Bulgaria, Canada, China, Colombia, Croatia, Cyprus, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, ISRAEL, Italy, JAPAN, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Maldives, Nepal, Netherlands, Norway, Oman, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovak Republic, Slovenia, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad & Tobago, Turkey, Uganda, Ukraine, United Arab Emirates.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 21, 2025
Recall initiation date
Apr 11, 2025
Center classification date
May 9, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records