Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syri... — FDA record Z-1726-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 27, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ANGIO PACK, Medline SKU # DYNJ57032D; 2. NEURO INTERVENTIONAL RADIOLOGY, Medline SKU # DYNJ64026C; 3. NEURO INERVENTIONAL RADIOLOGY, Medline SKU # VASC1650.
Reason
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Classification
Class I
Recalling firm
Medline Industries, LP
Quantity
2154 units
Agency-reported status
Ongoing
Source record identifier
Z-1726-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityMedline SKU # DYNJ57032D, UDI/DI each 10198459183850, UDI/DI case 40198459183851, Lot Number: 25HBN999; Medline SKU # DYNJ57032D, UDI/DI each 10198459183850, UDI/DI case 40198459183851, Lot Number: 25HBO039; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24JBC741; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24IBM270; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24IBE756; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24GBC735; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24FBF002; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24DBC238; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 24ABY360; Medline SKU # DYNJ64026C, UDI/DI each 10195327550059, UDI/DI case 40195327550050, Lot Number: 23LBH735; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 26ABP305; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 26ABI725; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25KBD893; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25HBM531; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25HBA055; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25EBC929; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25DBL096; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25DBD577; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25DBI493; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25ABI074; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 24LBG466; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 26BBG430; Medline SKU # VASC1650, UDI/DI each 10198459209666, UDI/DI case 40198459209667, Lot Number: 25LBC439.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 15, 2026
Recall initiation date
Feb 27, 2026
Center classification date
Apr 7, 2026