6.5 IN (17 cm) APPX 0.78 ml, EXT SET w/Clave CLEAR, Clave 4-WAY Stopcock, Spiros, RED CAP. Accessory for intravascula... — FDA record Z-1723-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 1, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
6.5 IN (17 cm) APPX 0.78 ml, EXT SET w/Clave CLEAR, Clave 4-WAY Stopcock, Spiros, RED CAP. Accessory for intravascular administration set.
Reason
Due to a manufacturing defect which results in a visible gap between the connector's top and bottom housing.
Classification
Class II
Recalling firm
ICU Medical, Inc.
Quantity
Total of all affected units = 5,795,989 units
Agency-reported status
Ongoing
Source record identifier
Z-1723-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct Number: CH3771 UDI Code: 0110887709098678172702013050105782919 Lot Numbers: 5782919

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Austria, Bahrain, Belgium, Canada, Colombia, Estonia, France, Germany, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Peru, Poland, Saudi Arabia, Slovakia, Slovenia, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam. O.U.S.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Sep 21, 2022
Recall initiation date
Aug 1, 2022
Center classification date
Sep 9, 2022