Washer Disinfector Aquadis 56; Model Number: 56A — FDA record Z-1720-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 2, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Washer Disinfector Aquadis 56; Model Number: 56A
Reason
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Classification
Class II
Recalling firm
Getinge Disinfection Ab
Quantity
102 units (8 US, 94 OUS)
Agency-reported status
Ongoing
Source record identifier
Z-1720-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel: 56A; UDI-DI: 07340153710344; Serial Numbers: WAA104857, WAA107934, WAA107937, WAA108359, WAA108540, WAA108963, WAA109338, WAA110115, WAA108201, WAA109175, WAA109884, WAA109989, WAA109998, WAA110804, WAA600035, WAA109641, WAA109651, WAA111118, WAA106820, WAA106835, WAA106841, WAA107693, WAA107428, WAA107429, WAA600055, WAA600060, WAA600061, WAA106607, WAA106668, WAA106753, WAA107358, WAA109804, WAA106793, WAA108419, WAA110679, WAA109022, WAA109030, WAA109564, WAA109566, WAA109595, WAA109598, WAA105283, WAA107346, WAA600188, WAA107314, WAA107330, WAA110062, WAA110075, WAA110084, WAA110104, WAA110163, WAA110302, WAA110201, WAA600558, WAA110953, WAA110963, WAA110138, WAA109975, WAA107281, WAA107303, WAA108972, WAA110228, WAA110162, WAA110727, WAA110738, WAA110742, WAA110803, WAA108669, WAA108670, WAA108673, WAA108722, WAA110762, WAA110631, WAA110636, WAA600167, WAA600168, WAA110482, WAA111159, WAA111170, WAA111183, WAA105284, WAA108615, WAA109514, WAA110174, WAA110178, WAA110898, WAA110907, WAA600077, WAA600177, WAA600338, WAA600339, WAA600340, WAA600341, WAA600342, WAA600343, WAA600344, WAA600394, WAA600705, WAA600706, WAA600711, WAA600712, WAA600713.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 14, 2025
Recall initiation date
Apr 2, 2025
Center classification date
May 5, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records