Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syri... — FDA record Z-1715-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 27, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medline SKU VASC1061; 2. CSTM MANIFOLD KIT - FRYE, Medline SKU VASC1074; 3. CRYO EP MANIFOLD KIT, Medline SKU VASC1198; 4. MANIFOLD KIT - 500PSI OFF, Medline SKU VASCSHPOFF1; 5. MANIFOLD KIT LOW PRESSURE ON, Medline SKU VASCSLPON1.
Reason
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Classification
Class I
Recalling firm
Medline Industries, LP
Quantity
1698 kits
Agency-reported status
Ongoing
Source record identifier
Z-1715-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityMedline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25IBN161; Medline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25GBF769; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25HBN798; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25GBL527; Medline SKU VASC1198, UDI/DI each 10193489664263, UDI/DI case 40193489664264, Lot Number 25GDA142; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25IBP989; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25GBL526; Medline SKU VASCSLPON1, UDI/DI each 10889942405510, UDI/DI case 40889942405511, Lot Number 25GBL070.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 15, 2026
Recall initiation date
Feb 27, 2026
Center classification date
Apr 7, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records