Recall archive / FDA device enforcement
Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Number... — FDA record Z-1715-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Apr 4, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Numbers: TFL-SLS, TFL-PLS Product Description: SOLTIVE Laser Systems use technology to increase procedural efficiency and flexibility to ureteroscopy, PCNL, and potential soft tissue applications. Component: No
- Reason
- Preset treatment parameters are not consistently being used in accordance with the IFU.
- Classification
- Class II
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 189
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1715-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | SOLTIVE Pro SuperPulsed Laser System: UDI-DI: TFL-SLS - 00821925044135; TFL-PLS - 00821925044111 All units included Powered Laser Surgical Instrument - Laser module component information Name: TFL Premium Laser Unit, TFL Standard Laser Unit UDI-DI: TFL-CPLU - 00821925044586; TFL-CSLU - 00821925044593 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide and the countries of AU, CA, SG, DE, JP, HK.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- May 7, 2025
- Recall initiation date
- Apr 4, 2025
- Center classification date
- May 1, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.