IceSeed 1.5 CX NEEDLE US, Cryoablation Needle, REF H7493967433170. The needles are intended to convert high-pressure ... — FDA record Z-1688-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 24, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
IceSeed 1.5 CX NEEDLE US, Cryoablation Needle, REF H7493967433170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
Reason
Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
Classification
Class II
Recalling firm
Boston Scientific Corporation
Quantity
244 units
Agency-reported status
Ongoing
Source record identifier
Z-1688-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN 00191506032692, Batch Numbers: 33744276, 33744277, 33744278, 33744600, 33931638, 33931639, 33931900, 33980480, 33980481, 33980482, 33989380, 33989381, 33989382, 34063066, 34063067, 34063068, 34063069, 34063070, 34084080, 34084081, 34084084, 34094111, 34094112, 34110333, 34110334, 34110335, 34110336, 34110337, 34168777, 34168778, 34204925, 34204926, 34204927, 34295709.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 7, 2025
Recall initiation date
Mar 24, 2025
Center classification date
Apr 29, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records