Intelli-C, Right side suspended Tabletop, REF: 03400000 — FDA record Z-1668-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 27, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Intelli-C, Right side suspended Tabletop, REF: 03400000
Reason
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
Classification
Class II
Recalling firm
NRT X-RAY A/S
Quantity
32 units
Agency-reported status
Ongoing
Source record identifier
Z-1668-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 05713464000527/Serial Numbers: 627004, 627007, 639001, 639002,639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 7, 2023
Recall initiation date
Apr 27, 2023