Recall archive / FDA device enforcement
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software v... — FDA record Z-1654-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Mar 18, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
- Reason
- When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
- Classification
- Class II
- Recalling firm
- Abbott Laboratories
- Quantity
- 19
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1654-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI 00380740017170 SNs: 35405BG 36034BG 36259BG 36326BG 36327BG 36328BG 36372BG 36545BG 36556BG 36583BG 36588BG 36589BG 54539BG 55361BG 55454BG 55520BG 55814BG 55828BG 72229BG |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Apr 30, 2025
- Recall initiation date
- Mar 18, 2025
- Center classification date
- Apr 22, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.