Recall archive / FDA device enforcement
Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,... — FDA record Z-1650-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Feb 10, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"
- Reason
- Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.
- Classification
- Class II
- Recalling firm
- Medline Industries, LP
- Quantity
- 106,621 US; 2,350 OUS
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1650-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | MSC8522 - Puracol UDI-DI 10080196285860 (each) 40080196285861 (case) Lot P2090 MSC8622EP - Puracol Plus UDI-DI 10080196296804 (each) 40080196296805 (case) Lot PL2107 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Domestic distribution nationwide US. International distribution to Mexico and Panama.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Apr 1, 2026
- Recall initiation date
- Feb 10, 2026
- Center classification date
- Mar 26, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.