InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA — FDA record Z-1650-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 4, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA
Reason
Insulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the cartridge holder; or the cartridge holder may be difficult to remove from the insulin pen. If the issue is present, and if the user does not have backup insulin they may experience temporary hyperglycemia.
Classification
Class II
Recalling firm
Medtronic MiniMed, Inc.
Quantity
8,986
Agency-reported status
Ongoing
Source record identifier
Z-1650-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF/UDI-DI(GTIN)/Lot: MMT-105NNBLNA/763000827182/D0020, D0034, D0036, D0038, D0045, D0046, D0050, D0057; MMT-105NNGYNA/763000827199/D0026, D0037, D0039, D0041, D0042, D0043, D0049; MMT-105NNPKNA/763000827205/D0029, D0030, D0040, D0051, D0062; MMT-105ELBLNA/763000827151/D0021, D0033, D0035, D0044, D0059, D0060, D0061; MMT-105ELGYNA/763000827168/D0031, D0032, D0047, D0068; MMT105ELPKNA/763000827175/D0027, D0028, D0048

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: OK, IA, NC, SC, MI, MN, CA, AZ, AL, MO, WA, IL, KS, TX, FL, MT, MD, OH, VA, NY, TN, HI, MA, NJ, PA, CO, WI, GA, CT, NE, OR, LA, NM, SD, ND, IN, NV, WV, RI, AR, NH

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 30, 2025
Recall initiation date
Mar 4, 2025
Center classification date
Apr 21, 2025