What the record says
- Product
- Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number (SMN)/Ref Catalog Number : 10464344/K6456
- Reason
- Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.
- Classification
- Class II
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 129 units US; 159 units OUS
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1648-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI: 0084276802852621340BB22120610464344840, 0084276802852622049BC23021810464344840, 0084276802852622112BF23042210464344840. Lot Numbers: lot 21340BB Exp Date: 2022-12-06; lot 22049BC (corrected lot number 9/8/2022) Exp Date: 2023-02-18; lot 22112BF Exp Date: 2023-04-20. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Austria, France, ¿¿ Germany, Italy, Switzerland.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Sep 7, 2022
- Recall initiation date
- Jul 18, 2022
- Center classification date
- Aug 29, 2022
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.