EarlyVue VS30 Vitals monitor with Software Versions A.00.02 and A.00.01 Product Number: 863380 — FDA record Z-1645-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 6, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
EarlyVue VS30 Vitals monitor with Software Versions A.00.02 and A.00.01 Product Number: 863380
Reason
EarlyVue VS30 Vital Signs Monitor and EarlyVue VS30 Vitals monitor Missing Calibration Alarm, a risk of an inaccurate CO2 measurement which may lead to a failure to recognize a change in patient condition.
Classification
Class II
Recalling firm
Philips North America Llc
Quantity
1421 units US; 3 units OUS
Agency-reported status
Ongoing
Source record identifier
Z-1645-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: (01)00884838091412 Software Version A.00.02 and A.00.01

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide Foreign: Belgium Brazil Canada Denmark Germany Guam Hong Kong Indonesia Italy Norway Poland Singapore Sweden Switzerland United Kingdom

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jun 7, 2023
Recall initiation date
Mar 6, 2023
Center classification date
May 26, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records