Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utili... — FDA record Z-1636-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 30, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).
Reason
The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.
Classification
Class II
Recalling firm
Tandem Diabetes Care, Inc.
Quantity
1,380 units
Agency-reported status
Ongoing
Source record identifier
Z-1636-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityVersion 7.8.1/ UDI: (01)008500066132625

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

International distribution in the countries of Czech Republic, Slovakia.

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Apr 1, 2026
Recall initiation date
Jan 30, 2026
Center classification date
Mar 25, 2026