To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural... — FDA record Z-1616-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Mar 31, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and the gastrointestinal tract Models: NA-U401SX-4022, NA-U401SX-4022-A
- Reason
- Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.
- Classification
- Class III
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 1297 (Boxes; 5 per Box)
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1616-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI: 04953170388286 Lot numbers: KR248740 KR248746 KR252863 KR253666 KR253681 KR260854 KR260865 KR260873 KR260875 KR260876 KR260877 KR260896 KR260917 KR260937 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Natiowide Foreign: Canada, Germany
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- May 31, 2023
- Recall initiation date
- Mar 31, 2023
- Center classification date
- May 19, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.