ViziShot 2 Single Use Aspiration Needle, 21G- To be used with ultrasound endoscopes for ultrasound guided fine needle... — FDA record Z-1615-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 31, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ViziShot 2 Single Use Aspiration Needle, 21G- To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and the gastrointestinal tract. Model: NA-U401SX-4021, NA-U401SX-4021-A
Reason
Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.
Classification
Class III
Recalling firm
Olympus Corporation of the Americas
Quantity
2058 (Boxes; 5 per Box)
Agency-reported status
Ongoing
Source record identifier
Z-1615-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 04953170388248 Lot numbers: KR243729 KR248648 KR248652 KR248673 KR248784 KR248785 KR248808 KR249160 KR251593 KR257313 KR257314 KR264662 KR264697 KR264708 KR264711 KR264726 KR264732 KR264746 KR264751 KR264778 KR264806 KR276770

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Natiowide Foreign: Canada, Germany

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
May 31, 2023
Recall initiation date
Mar 31, 2023
Center classification date
May 19, 2023