ET III Nano-HA fixture System ET NH Dental Implant Model # ET3R5010B -Intended for Dental Implant Catalog #: AET3R5010B — FDA record Z-1604-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 6, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ET III Nano-HA fixture System ET NH Dental Implant Model # ET3R5010B -Intended for Dental Implant Catalog #: AET3R5010B
Reason
Mis-labeled length of the dental implant on the Recall Implant's package. Dental implant size D5.0 x L8.5mm, labeled as D5.0 x L10mm
Classification
Class II
Recalling firm
Hiossen Inc.
Quantity
287 units
Agency-reported status
Ongoing
Source record identifier
Z-1604-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00814913023717 Lot Number: H1E22J263

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 24, 2023
Recall initiation date
Apr 6, 2023
Center classification date
May 18, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.