Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- ... — FDA record Z-1598-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Mar 21, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355 Artis zee floor MN- 10094142 Artis zee biplane MN- 10094143
- Reason
- Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
- Classification
- Class II
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 583 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1598-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Model/UDI-DI: Artis zee floor- 4056869010045 Artis zee ceiling- 4056869010052 Artis zee multi-purpose- 4056869010076 Artis zee biplane- 4056869010069 Artis zee floor MN- 4056869010090 Artis zee biplane MN- 4056869010106 Artis zeego- 4056869010083 Artis Q floor- 4056869009971 Artis Q ceiling- 4056869009988 Artis Q biplane- 4056869009995 Artis Q zeego- 4056869010007 Artis Q.zen floor- 4056869010014 Artis Q.zen ceiling- 4056869010021 Artis Q.zen biplane- 4056869010038 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Apr 24, 2024
- Recall initiation date
- Mar 21, 2024
- Center classification date
- Apr 18, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.