RENASYS EDGE 300ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pump (NPWT), as ... — FDA record Z-1596-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 11, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
RENASYS EDGE 300ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pump (NPWT), as it may promote wound healing. Product Number: 66803139
Reason
Black particulate matter from the carbon filter component of the canisters entering the NPWT pump and potentially causing damage to the pump
Classification
Class II
Recalling firm
Smith & Nephew Medical, Ltd.
Quantity
6895 units
Agency-reported status
Ongoing
Source record identifier
Z-1596-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: (01)05000223501044 Lot Numbers: 1016448, 1016478, 1016537, 1016773, 1017147, 1017417, 1017418, 1017912, 1016044

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Chile.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 24, 2024
Recall initiation date
Mar 11, 2024
Center classification date
Apr 18, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.