SW Bari Lift & Transfer, Model Number 38060000 — FDA record Z-1595-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 25, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
SW Bari Lift & Transfer, Model Number 38060000
Reason
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
Classification
Class II
Recalling firm
Raye's Inc.
Quantity
1,149 units
Agency-reported status
Ongoing
Source record identifier
Z-1595-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 00845699001040, All serial numbers

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, UK.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 24, 2024
Recall initiation date
Jan 25, 2024
Center classification date
Apr 18, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.