Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Softwa... — FDA record Z-1573-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 18, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package
Reason
Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
Classification
Class II
Recalling firm
Baxter Healthcare Corporation
Quantity
59
Agency-reported status
Ongoing
Source record identifier
Z-1573-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Code: " Product Code: HKO " Version or Model: 901029 " Catalog Number: VS100 " Company Name: WELCH ALLYN, INC. " Primary DI Number: 00732094214154 " Packaging DI Number: 00732094212228 " Device Description: VS100 UDI BASE UNIT " Software version: 3.2.0.1 " Serial Numbers: Refer to attached Affected Product Table

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AZ, CO, IL, IN, KS, MA, NE, NJ, NM, NY, PA, SC, TX, VA and the countries of Colombia and India.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 25, 2026
Recall initiation date
Feb 18, 2026
Center classification date
Mar 18, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.