Aesculap Caiman Articulating D5/360MM - An electrosurgical cutting and coagulation device for cutting/coagulation/abl... — FDA record Z-1569-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 21, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Aesculap Caiman Articulating D5/360MM - An electrosurgical cutting and coagulation device for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery, and may in addition be intended for open surgery. Product Code: PL771SU
Reason
Potential for the spring to become loose and fall out of the device, if the spring falls out during a surgical procedure, it may enter the patient's body.
Classification
Class II
Recalling firm
Aesculap Implant Systems LLC
Quantity
6 pieces
Agency-reported status
Ongoing
Source record identifier
Z-1569-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 04046964837340 Lot Number: 52785146

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of OH.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 17, 2023
Recall initiation date
Mar 21, 2023
Center classification date
May 5, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.