Welch Allyn CONNEX Accessory Power Management Stand: — FDA record Z-1559-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 24, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Welch Allyn CONNEX Accessory Power Management Stand:
Reason
There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Classification
Class II
Recalling firm
Baxter Healthcare Corporation
Quantity
114,202 units
Agency-reported status
Ongoing
Source record identifier
Z-1559-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll Serial Numbers. Product Code: 7000-APM, UDI-DI: 00732094210613; Product Code: 75-HCA-CTB, UDI-DI: 00732094240597; Product Code: 75-HCA-MTB, UDI-DI: 00732094240603;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 23, 2025
Recall initiation date
Mar 24, 2025
Center classification date
Apr 14, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.