Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and so... — FDA record Z-1549-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 13, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using the Atellica¿ CI Analyzer and Atellica¿ CH Analyzer Siemens Material Number (SMN): 11537211
Reason
Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts calibrator, quality control (QC), and patient results.
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
8721 units
Agency-reported status
Ongoing
Source record identifier
Z-1549-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00630414610849 All lots

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Curacao, St¿Eus, Czech Republic, Denmark, Estonia, Finland, France, Germany, ¿ Greece, Hungary, India, Ireland, Israel, Italy, Latvia, Lithuania, Mexico, Netherlands, ¿¿¿¿ New Zealand, Norway, Paraguay, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., United¿Kingdom.¿

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 24, 2024
Recall initiation date
Mar 13, 2024
Center classification date
Apr 16, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records