Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Descripti... — FDA record Z-1547-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 6, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR Kit/REA4011-01, Meijer TRUE METRIX AIR Kit/REA4019-01, Publix TRUE METRIX AIR Kit/REA4020-01, Discount Drug Mart TRUE METRIX AIR Kit/REA4025-01, HEB TRUE METRIX AIR Kit/REA4031-01, SunmarkTRUE METRIX AIR Kit/REA4054-01, Rite AidTRUE METRIX AIR Kit/REA4066-01, Centerwell TRUE METRIX AIR Kit/REA4081-01, HumanaTRUE METRIX AIR Meter Only/REA4081-40, Centerwell TRUE METRIX AIR Starter Kit/REA4081-43, Healthmart TRUE METRIX AIR Kit/REA4089-01, Relion TRUE METRIX AIR Kit/REA4094-01, Relion TRUE METRIX AIR Kit/REA4094-01RX, Kroger TRUE METRIX AIR Kit/REA4203-01, Foster and Thrive TRUE METRIX AIR Kit/REA4211-01, TRUE METRIX AIR Kit/REA4H01-01, TRUE METRIX AIR NFRS Meter Only/REA4H01-40, TRUE METRIX AIR NFRS Starter Kit/REA4H01-43, TRUE METRIX AIR Kit (mg/dL)/REA4i03-01, TRUE METRIX AIR Starter Kit (mg/dL)/REA4i03-02, Walmart (MediMart) TRUE METRIX AIR Meter Only (Mexico - mg/dL)/REA4i04-00, TRUE METRIX AIR Kit (Australia - mmol/L)/REA4i81-11, TRUE METRIX AIR Kit (United Kingdom - mmol/L)/REA4i82-11, TRUE METRIX AIR Starter Kit (United Kingdom - mmol/L)/REA4i82-12
Reason
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
Classification
Class I
Recalling firm
Trividia Health, Inc.
Quantity
3,678,026
Agency-reported status
Ongoing
Source record identifier
Z-1547-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityDevice Item/UDI-DI: REA4001-01/311917173894, REA4002-01/96295129267, REA4007-01/50428354766, REA4011-01/87701427398, REA4019-01/708820685659, REA4020-01/41415072787, REA4025-01/93351022784, REA4031-01/41220009169, REA4054-01/10939742445, REA4066-01/11822590051, REA4081-01/21292015015, REA4081-40/21292007164, REA4081-43/21292007584, REA4089-01/52569139042, REA4094-01/681131240154, REA4094-01RX/681131403214, REA4203-01/41260007354, REA4211-01/10939958617, REA4H01-01/21292007010, REA4H01-40/21292007447, REA4H01-43/21292007911, International: REA4i03-01/N/A, REA4i03-02/N/A, REA4i04-00/21292012731, REA4i81-11/N/A, REA4i82-11/21292012212, REA4i82-12/21292009991. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX AIR Owner's Booklet/ REA4TVH03 Rev 57 and prior, TRUE METRIX AIR Owner's Booklet (English)/ REA4TVH35 Rev 51 and prior, TRUE METRIX AIR Owner's Booklet (Spanish)/ REA4TVH35S Rev 51 and prior, Relion (Walmart) TRUE METRIX AIR Owners Booklet/ REA4RLN03 Rev 52 and prior, Relion (Walmart) TRUE METRIX AIR Owner's Booklet - English/ REA4RLN35 Rev 52 and prior, TRUE METRIX AIR Owner's Booklet International Trividia House Brand/ REA4ITV03 Rev 55 and prior, TRUE METRIX AIR Owners Booklet for Trividia Health UK Limited/ REA4UKT03 Rev 55 and prior, Trividia Health LATAM - TRUE METRIX AIR Owners Booklet/ REA4LAT03 Rev 51 and prior, TRUE METRIX AIR Owners Booklet for Trividia Health Australia/ REA4AUT03 Rev 58 and prior. Device serial numbers: Prefixes "TB" or "TA" followed by a seven-digit numeric sequence.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 18, 2026
Recall initiation date
Feb 6, 2026
Center classification date
Mar 12, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.