Recall archive / FDA device enforcement
Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 3010E — FDA record Z-1542-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 19, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 3010E
- Reason
- Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
- Classification
- Class II
- Recalling firm
- Smiths Medical ASD Inc.
- Quantity
- 27 pumps
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1542-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | All serial numbers for pumps manufactured or serviced with these potentially affected Barrel Clamp Guides between Jul-2016 and Apr-2021. a. 3010, UDI/DI none, Serial Numbers: M03315, M03910; b. 3010A, UDI/DI 10610586032097, Serial Numbers: M04601, M05551, M05782, M06829, M07259, M07334, M07339, M07342, M07345, M07371, M08269, M08750, M08815, M10253, M10267, M10871, M10876, M10965, M11872, M11896, M11899, M11901, M11905, M11915; c. 3010E, UDI/DI none, Serial Numbers: M05179 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Apr 24, 2024
- Recall initiation date
- Dec 19, 2023
- Center classification date
- Apr 15, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.