stryker Blueprint Mixed Reality Glenoid Digitizer, Catalog Number MRUE201, packaged in stryker Blueprint Mixed Realit... — FDA record Z-1535-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 5, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
stryker Blueprint Mixed Reality Glenoid Digitizer, Catalog Number MRUE201, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number MRUEGLN; Total Shoulder Arthroplasty
Reason
Stryker identified an elevated complaint rate related to difficulties using the Blueprint Mixed Reality Guidance Instrumentation system is multifactorial, but in each instance prohibits the system from meeting its intended use. The most commonly reported complaints were related to impaired software functioning, impaired instrumentation functioning, and general difficulties using the product.
Classification
Class II
Recalling firm
Tornier S.A.S.
Quantity
22 units
Agency-reported status
Ongoing
Source record identifier
Z-1535-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll lots manufactured through Oct 2024, Distributed 23-Aug-2023 through 31-Dec-2024, GTIN 03700434022933

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: TX, MN, AK, PA, MA, TN, ID, and France, Canada, and Ireland

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 16, 2025
Recall initiation date
Mar 5, 2025
Center classification date
Apr 8, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records