ARIES SARS-CoV-2 Assay — FDA record Z-1524-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jul 14, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ARIES SARS-CoV-2 Assay
Reason
SARS-CoV-2 assay that detects SARS-CoV-2 ORF1ab and N genes may not detect the ORF1ab gene. A positive results if either SARS-CoV-2 genes is detected, but If both genes are mistakenly not detected, a false negative result would be generated.
Classification
Class II
Recalling firm
Luminex Corporation
Quantity
257
Agency-reported status
Ongoing
Source record identifier
Z-1524-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot: AB4042A

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S. Nationwide distribution in the states of NE, TX, MS, TN, KY, ME, CA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 24, 2022
Recall initiation date
Jul 14, 2022
Center classification date
Aug 17, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.