The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL... — FDA record Z-1508-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 24, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX¿ cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid Profile Cassette/10-989; Cholestech LDX TC HDL Cassette/10-987
Reason
Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
Classification
Class II
Recalling firm
Alere San Diego, Inc.
Quantity
436,990
Agency-reported status
Ongoing
Source record identifier
Z-1508-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll lots of Part/UDI-DI: 10-991/00893038002654, 99021/00899722002344, 14-531/04571226470831, 10-990/00893038002647, 10-989/00893038002630, 10-987/00893038002616

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR, HK, BR, CO, AR, CA

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 17, 2022
Recall initiation date
Jun 24, 2022
Center classification date
Aug 6, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records