Recall archive / FDA device enforcement
HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended U... — FDA record Z-1506-2022
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jun 10, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- HAIOU Needle retractable safety syringe with needle-(1ml syringe with 25G x 1-inch needle} 200 pieces/box. Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-25G25
- Reason
- Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user
- Classification
- Class II
- Recalling firm
- Guangdong Haiou Medical Apparatus Co., Ltd
- Quantity
- 4,050,000 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1506-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Lot Number: 20IB0 Exp. Date: 2025/09/04 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the state of CA.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Aug 17, 2022
- Recall initiation date
- Jun 10, 2022
- Center classification date
- Aug 5, 2022