What the record says

Product
ADVIA Centaur Cortisol 50T (Material Number 10994924), ADVIA Centaur Cortisol 250T (Material Number 10994926), and ADVIA Centaur Cortisol REF 250T (Material Number 10994927). In vitro diagnostic use in the quantitative determination of cortisol in human serum, plasma (EDTA and lithium heparin), and urine.
Reason
There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
18,961 kits
Agency-reported status
Ongoing
Source record identifier
Z-1505-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityADVIA Centaur Cortisol (50T) UDI-DI: 00630414602943; Kit Lot Numbers: 49741344 07241346 88318346 29109348 and higher; ADVIA Centaur Cortisol 250T UDI-DI: 00630414602950; Kit Lot Numbers: 49742344 88319346 29108348 and higher; ADVIA Centaur Cortisol (REF) 250T UDI-DI: 00630414598659; Kit Lot Numbers: 75474344 11352346 and higher

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 10, 2023
Recall initiation date
Mar 16, 2023
Center classification date
May 3, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.