Atellica IM Cortisol 50T (Material Number 10995538), 250T (Material Number 10995537), and REF 250T (Material Number 1... — FDA record Z-1504-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 16, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Atellica IM Cortisol 50T (Material Number 10995538), 250T (Material Number 10995537), and REF 250T (Material Number 11206248). In vitro diagnostic use in the quantitative determination of cortisol in human serum, plasma (EDTA and lithium heparin), and urine.
Reason
There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
34,746 kits
Agency-reported status
Ongoing
Source record identifier
Z-1504-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAtellica IM Cortisol (50T) UDI-DI: 00630414598659; Kit Lot Numbers: 50732343 50733343 88975345 29106347 29107347 and higher; Atellica IM Cortisol 250T UDI-DI: 00630414598642; Kit Lot Numbers: 50734343 88974345 29105347 and higher; Atellica IM Cortisol (REF) 250T UDI-DI: 00630414293608; Kit Lot Numbers: 53842343 58039343 94430345 and higher

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 10, 2023
Recall initiation date
Mar 16, 2023
Center classification date
May 3, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.