Phoroptor VRx Digital Refraction System Model Numbers: 16242 — FDA record Z-1501-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 28, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Phoroptor VRx Digital Refraction System Model Numbers: 16242
Reason
The head of the phoropter head could come loose and potentially detach due to a default in the assembly.
Classification
Class II
Recalling firm
Reichert, Inc.
Quantity
1876 units
Agency-reported status
Ongoing
Source record identifier
Z-1501-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel Numbers: 16242; UDI-DI: 00812559011730; Serial Numbers: TBD; Note: The majority of units within the scope of the planned recall are older devices which do not have a UDI on the label.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic: Nationwide Distribution. International: Austria, Bolivia, Canada, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Morocco, Netherlands, Norway, Qatar, Saudi Arabia, Singapore, Slovakia, Switzerland, Thailand, United Arab Emirates, and United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 9, 2025
Recall initiation date
Feb 28, 2025
Center classification date
Mar 31, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.