IBA Proton Therapy System - PROTEUS 235 — FDA record Z-1497-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 3, 2026. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
IBA Proton Therapy System - PROTEUS 235
Reason
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Classification
Class II
Recalling firm
Ion Beam Applications S.A.
Quantity
4 units (1 US, 3 OUS)
Agency-reported status
Ongoing
Source record identifier
Z-1497-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 18, 2026
Recall initiation date
Feb 3, 2026
Center classification date
Mar 6, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.