CryoTreQ- Aa disposable handheld instrument intended for ophthalmic surgery. It creates a tip at cryogenic temperatur... — FDA record Z-1497-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 17, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
CryoTreQ- Aa disposable handheld instrument intended for ophthalmic surgery. It creates a tip at cryogenic temperatures by evaporation of N2O that can be utilized to perform cryotherapy based on the destruction of tissue by extreme cold to locally perform cryocoagulation on tissue. Ref: CT00.D01
Reason
Device suddenly and forcefully disassembled after cryotherapy, potential for long-term or permanent impairment of a patient or user if a CryoTreq device suddenly or forcefully disassembles during a cryosurgery procedure.
Classification
Class II
Recalling firm
Beaver Visitec International, Inc.
Quantity
501 (USA)
Agency-reported status
Ongoing
Source record identifier
Z-1497-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 08719214223458 All Affected Lots: 20214450, 20214366, 20214367, 20214518, 20214522, 20214615, 20214744, 20214745, 20214961, 20214962, 20215203, 20215204, 20215479, 20216263, 20216264, 20216010, 20227952, 20227950, 20227951, 20228408, 20228657, 20228897, 20228898, 20229034, 20229138, 20229334, 20229460, 20229630, 20229748, 20230261, 20221083, 20221140, 20221364, 20221746, 20221894, 20222365

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 10, 2023
Recall initiation date
Mar 17, 2023
Center classification date
May 2, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.