LUMINOS Lotus Max (VF11) — FDA record Z-1481-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 23, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
LUMINOS Lotus Max (VF11)
Reason
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
704 units in the United States (1945 units worldwide)
Agency-reported status
Ongoing
Source record identifier
Z-1481-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel: 11574100

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide - Worldwide Distribution

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 31, 2023
Recall initiation date
Nov 23, 2022
Center classification date
May 23, 2023