Recall archive / FDA device enforcement
Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Num... — FDA record Z-1479-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Feb 25, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITL
- Reason
- There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
- Classification
- Class II
- Recalling firm
- Stryker Corporation
- Quantity
- 182,344 total units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1479-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | 5400-030-008 UDI-DI: 07613154340842, Lot Numbers: 2023012515 2023012715 2023022012 2023050814 2023072632 2023102312 2023102612 2024032514 2024040212; 5400-030-008ITL UDI-DI: 07613327359473, Lot Numbers: 2022092714 2023020633 2023032433 2023050915 2023052514 2023062216 2023082814 2023101112 2024031812 2022092914 2023020814 2023032814 2023051114 2023053115 2023062032 2023090618 2023101714 2024042212 2022101332 2023020915 2023033016 2023051215 2023060914 2023082215 2023100312 2023102014 2024042312 2022121932 2023021416 2023050115 2023052214 2023061412 2023082312 2023100412 2024030414 2024061814 2022122032 2023022032 2023050515 2023052412 2023061514 2023082514 2023100912 2024030512 2024062814 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Apr 9, 2025
- Recall initiation date
- Feb 25, 2025
- Center classification date
- Mar 28, 2025