SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200 — FDA record Z-1474-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 2, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200
Reason
Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.
Classification
Class II
Recalling firm
Stryker Corporation
Quantity
5,228
Agency-reported status
Ongoing
Source record identifier
Z-1474-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityRef/UDI-DI/Lot: 0233-050-100/07613327051650/159190, 159191, 159192, 159194, 159195, 159196, 159197, 159198, 159199, 159200, 159201, 159202, 159203, 159204, 159205, 159206, 159207, 159208, 159209, 159210, 159211, 159212, 159213, 159214, 159215, 159216, 159217, 159218, 164764, 164765, 164766, 164767, 164768, 164769, 164770, 164771, 164772, 164773, 164774, 164775, 164776, 164777, 164778, 164779, 164780, 164781, 164782, 164783, 164784, 164785, 164786, 164787, 164788, 164789, 164790, 164791, 164792, 164793, 164794, 164795, 164796, 164797, 164798, 164799, 164800, 164801, 164802, 164803, 164804, 164805, 164806, 164807, 164808, 164809, 164810, 164811, 164812, 164813, 164814, 164815, 164816, 164817, 164818, 164819, 164820, 164821, 164822, 168940, 168941, 168942, 168943, 168944, 168945, 168946, 168947, 168948, 168949, 168950, 168951, 168952, 168953, 168954, 168955, 168956, 168957, 168958, 168959, 168960, 168961, 168962, 168963, 168964, 168965, 168966, 168967, 168968, 168969, 168970, 168971, 168972, 168973, 168974, 168975, 168976, 168977, 168978, 168979, 168980, 168981, 168987; 0233-050-200/07613327051643/158865, 158868, 158869, 158872, 158874, 158875, 158878, 158884, 165446, 165448, 165451, 165457, 165458, 165459, 165461, 165462, 165464, 168909, 168910, 168911, 168912, 168913, 168914, 168915, 168916, 168917, 168918, 168919, 168920, 168921, 168923, 168924, 168925, 168926, 168927, 168928

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of PA, MS, IL, OH, CA, TX, MO, IN, KS, OK, MT, MA, NE, AZ, NV, NJ, IA, WV, MD, KY, FL, NM, NY, CO, WA, NC, VA, LA, ND, WY, HI SC, UT, OR, SD, AR, TN, AK, CT, GA, ID, VT, NH, MN, WI, DE, RI, AL, MI and the countries of NL, PE, AT, CA, TW, PH, SG, KR, CN.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 10, 2023
Recall initiation date
Mar 2, 2023
Center classification date
Apr 28, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records