VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b... — FDA record Z-1464-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 14, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b) AST-GN95, REF 421982; c) AST-GN99, REF 423102; d) AST-N390, REF 423340; e) AST-N391, REF 423341; f) AST-N392, REF 423342; g) AST-N395, REF 423491; h) AST-N401, REF 423643; i) AST-N402, REF 423644; j) AST-N404, REF 423664; k) AST-N802, REF 423706; l) AST-N405, REF 423864; m) AST-N417, REF 423880; n) AST-N408, REF 423924; o) AST-N409, REF 423925; p) AST-N420, REF 424039; q) AST-N423, REF 424042; r) AST-N422, REF 424056; s) AST-N436, REF 424440; t) AST-N809, REF 424703; u) AST-N806, REF 424709; v) AST-N807, REF 424710; w) AST-N808, REF 424711; x) AST-N810, REF 424712; y) AST-N812, REF 424721
Reason
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.
Classification
Class I
Recalling firm
Biomerieux Inc
Quantity
49215 units
Agency-reported status
Ongoing
Source record identifier
Z-1464-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitya) AST-N334, REF 418984, UDI/DI 3573026490720, Lot/Serial Numbers: 7742790503, 7742805403, 7742828503, 7742830203; b) AST-GN95, REF 421982, UDI/DI 3573026565923, Lot/Serial Numbers: 6852740103, 6852751503, 6852778503, 6852811503; c) AST-GN99, REF 423102, UDI/DI 3573026608132, Lot/Serial Numbers: 6892721103, 6892727403, 6892740503, 6892740403, 6892743503, 6892761503, 6892770403, 6892790103, 6892799403, 6892799103, 6892810103, 6892818103, 6892831503; d) AST-N390, REF 423340, UDI/DI 3573026612177, Lot/Serial Numbers: 1402832104, 1402761504; e) AST-N391, REF 423341, UDI/DI 3573026612269, Lot/Serial Numbers: 1412831104, 1412754504; f) AST-N392, REF 423342, UDI/DI 3573026612283, Lot/Serial Numbers: 1422819204; g) AST-N395, REF 423491, UDI/DI 3573026615802, Lot/Serial Numbers: 1452725204; h) AST-N401, REF 423643, UDI/DI 3573026616816, Lot/Serial Numbers: 1512776404, 1512789204, 1512811104, 1512831104, 1512720404, 1512729404, 1512729104, 1512748504, 1512754104, 1512785404; i) AST-N402, REF 423644, UDI/DI 3573026616830, Lot/Serial Numbers: 1522720504, 1522733404; j) AST-N404, REF 423664, UDI/DI 3573026617356, Lot/Serial Numbers: 1542819204; k) AST-N802, REF 423706, UDI/DI 3573026617752, Lot/Serial Numbers: 0422750104, 0422763104, 0422764504, 0422782404, 0422820504, 0422820404; l) AST-N405, REF 423864, UDI/DI 3573026621360, Lot/Serial Numbers: 1552757504, 1552792404, 1552800404, 1552820104, 1552733104, 1552741504, 1552757404; m) AST-N417, REF 423880, UDI/DI 3573026621704, Lot/Serial Numbers: 0192763504, 0192819404; n) AST-N408, REF 423924, UDI/DI 3573026622725, Lot/Serial Numbers: 1582810504, 1582820104, 1582734404, 1582742504, 1582764404, 1582782504; o) AST-N409, REF 423925, UDI/DI 3573026622749, Lot/Serial Numbers: 1592811104, 1592746404; p) AST-N420, REF 424039, UDI/DI 3573026625283, Lot/Serial Numbers: 0462810404; q) AST-N423, REF 424042, UDI/DI 3573026625337, Lot/Serial Numbers: 0502750404, 0502825504; r) AST-N422, REF 424056, UDI/DI 3573026625610, Lot/Serial Numbers: 0482819504; s) AST-N436, REF 424440, UDI/DI 3573026631666, Lot/Serial Numbers: 0672734104, 0672754104; t) AST-N809, REF 424703, UDI/DI 3573026635978, Lot/Serial Numbers: 0962799404; u) AST-N806, REF 424709, UDI/DI 3573026636067, Lot/Serial Numbers: 0932729404; v) AST-N807, REF 424710, UDI/DI 3573026636081, Lot/Serial Numbers: 0942723404; w) AST-N808, REF 424711, UDI/DI 3573026636104, Lot/Serial Numbers: 0952758104; x) AST-N810, REF 424712, UDI/DI 3573026636128, Lot/Serial Numbers: 0972729504; y) AST-N812, REF 424721, UDI/DI 3573026636296, Lot/Serial Numbers: 0992777104, 0992776504, 0992821404

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
May 1, 2024
Recall initiation date
Mar 14, 2024
Center classification date
Apr 25, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records