Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-FLEX7.5 (G576... — FDA record Z-1427-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 7, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-FLEX7.5 (G57695), C-PTIS-100-HC-G-EU-FLEX8.5 (G57696), C-PTIS-100-HC-G-NA-FLEX7.5 (G57691), C-PTIS-100-HC-G-NA-FLEX8.5 (G57692)
Reason
There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.
Classification
Class I
Recalling firm
Cook Incorporated
Quantity
3.053 units
Agency-reported status
Ongoing
Source record identifier
Z-1427-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityC-PTIS-100-HC-G-EU-FLEX7.5 (G57695): UDI-DI 00827002576957, Lot Number 14842940; C-PTIS-100-HC-G-EU-FLEX8.5 (G57696): UDI-DI 00827002576964, Lot Numbers 14727192, 14332964, 14335896, 14380758, 14380755, 14332978, 14332973, 14335892, 14335893, 14464000, 14332985, 14464008, 14463988, 14503228, 14503224, 14503222, 14503229, 14503227, 14742869, 14742870, 14727207, 14742873; C-PTIS-100-HC-G-NA-FLEX7.5 (G57691): UDI-DI 00827002576919, Lot Numbers 15063922, 15063923, 15103482, 14227740, 14221180, 14204927, 14227743, 14221125, 14221198, 14221190, 14163328, 14250265, 14250269, 14317245, 14317213, 14345634, 14376893, 14633088, 14633089, 14794549, 14794548, 14794547, 14874486, 14874484, 14926500, 14967737, 14926505, 14967739, 14967741, 14221180; C-PTIS-100-HC-G-NA-FLEX8.5 (G57692): UDI-DI 00827002576926, Lot Numbers 14608362, 14750607, 14894816, 14264399, 14299799, 14299802, 14299803, 14345667, 14365693, 14365699, 14365696, 14365694, 14365701, 14345687, 14345587, 14456438, 14456443, 14456431, 14407846, 14750610, 14488394, 14488418, 14608361, 14608363, 14750618, 14608360, 14961742, 14961716, 15089037, 15089038, 15136113

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 17, 2023
Recall initiation date
Apr 7, 2023
Center classification date
May 10, 2023