What the record says

Product
Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0
Reason
Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range; Cholesterol (CHOL) Instrument Specific Application using the recommended calibration material CAL2351, the product fails to meet its linearity performance claim quoted on the kit inserts. The linearity is reduced by up to -35% however Quality Control results will still fall within assigned ranges. This may lead to delayed results.
Classification
Class II
Recalling firm
Randox Laboratories Ltd.
Quantity
15 kits
Agency-reported status
Terminated
Source record identifier
Z-1424-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN : 05055273201130 Lot Number: 586177 Expiry Date: 28th Jan 2024

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide Distribution - MD, OR, PR, MT, WV

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 27, 2022
Recall initiation date
Jun 22, 2022
Center classification date
Jul 20, 2022
Termination date
May 10, 2024