Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model... — FDA record Z-1403-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 3, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627
Reason
Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.
Classification
Class II
Recalling firm
ZIMMER ORTHOPEDICS MANFACTURING LIMITED-GALWAY
Quantity
90 US; 136 OUS
Agency-reported status
Ongoing
Source record identifier
Z-1403-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityItem No. 113605, UDI-DI 00887868267550 Lots 66173214, 66173215; Item No. 113606, UDI-DI 00887868267567 Lots 66173216, 66173218, 66173219; Item No. 113607, UDI-DI 00887868267574 Lots 66173220, 66173221; Item No. 113627, UDI-DI 00887868267741 Lots 65901697, 65909480, 65915809

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US distribution nationwide. International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 26, 2025
Recall initiation date
Mar 3, 2025
Center classification date
Mar 20, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records