CereLink ICP Monitor; Catalog No. 826820, 826820P. Intracranial pressure monitor. — FDA record Z-1395-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 22, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
CereLink ICP Monitor; Catalog No. 826820, 826820P. Intracranial pressure monitor.
Reason
Integra has received complaints associated with ICP readings drifting to -50 mmHg (out-of-range) and may manifest an error message "sensor or extension cable failure!" The firm has found that this is caused by electrical interference from a component of the monitor's circuit board and from the environment.
Classification
Class I
Recalling firm
Integra LifeSciences Corp.
Quantity
1,210 units
Agency-reported status
Ongoing
Source record identifier
Z-1395-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI- 10381780533778; All serial numbers.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide Distribution: US nationwide, Australia, New Zealand, Canada, Hong Kong, Pakistan, Georgia, Saudi Arabia, Israel, South Africa, United Arab Emirates, Austria, Belgium, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Portugal, Switzerland, Spain, Sweden, Slovakia, and United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Aug 3, 2022
Recall initiation date
Jun 22, 2022
Center classification date
Jul 22, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.