Endo Model Modular Rotational Tibia - Small (Model No #15-2814/02), Medium (Model No #15-2814/03), and Large (Model N... — FDA record Z-1377-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 24, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Endo Model Modular Rotational Tibia - Small (Model No #15-2814/02), Medium (Model No #15-2814/03), and Large (Model No #15-2814/04). orthopedic prosthesis.
Reason
There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
Classification
Class II
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Quantity
3785
Agency-reported status
Ongoing
Source record identifier
Z-1377-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI 04026575316281 (Small) 04026575316298 (Medium) 04026575316304 (Large) Serial/Lot Numbers: 210420/0762 210426/0311 210426/0363 210426/0364 210426/0374 210504/2485 210504/2552 210504/2564 210504/2566 210510/2845 210510/2846 210602/0215 210602/0265 210602/0266 210602/1068 210602/2717 210602/2721 210602/2733 210602/2769 210629/0893 210629/0919 210629/0929 210629/0967 210629/0969 210629/1046 210629/1051 210629/1054 210629/2750 210802/0663 210802/0771 210802/0841 210907/2316 210907/2317 210913/0613 210913/0622 210913/0762 210913/0797 210913/0801 210913/0809 210913/2167 211025/0391 211025/2541 211025/2543 211025/2563 211102/0426 211102/0431 211115/0220 211115/1948 211115/2111

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN, TX, WA. Foreign distribution to Algeria Argentina Australia Austria Belgium Bulgaria Canada China Colombia Denmark Ecuador Estonia Finland France Germany Great Britain Greece Hungary India Indonesia Israel Italy Latvia Libya Lithuania Mexico Netherlands Norway Pakistan Peru Philippines Poland Romania Saudi Arabia Slovakia Slovenia Spain Sudan Sweden Switzerland Thailand Turkey Ukraine United Arab Emirates Uruguy Vietnam

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 20, 2022
Recall initiation date
May 24, 2022
Center classification date
Jul 12, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.