BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: ... — FDA record Z-1375-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 6, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES Desktop PC, REF: 1156-00, CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01, CII Safe ES Desktop PC, W-less SCNR, REF: 107-255-01, Used with Enterprise Server (ES), REF: 1115-00 with affected Enterprise Server software.
Reason
Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.
Classification
Class II
Recalling firm
CareFusion 303, Inc.
Quantity
727
Agency-reported status
Ongoing
Source record identifier
Z-1375-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityBD Pyxis Enterprise Server software versions: 1.7.4.128, 1.7.4.134, 1.7.4.136, 1.7.4.137, All 1.7.1, All 1.7.2, All 1.7.3. REF/UDI-DI: 327/10885403477836, 323/10885403512667, 352/10885403512674, 1116-00/10885403512605

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of KY, OH, NY, MO, NC, PA, FL, TN, CT, LA, MN, VA, MD, CA, TX, NE, IA, OK, MI, PR, SD, IN, ME, KS, IL, NM, WA, AR, HI, MS, GA, WY, WI, VT, AL, NV, NJ, DC, AZ, MT, AK, OR, CO, NH, MA, SC, ID, UT, ND and the countries of TW, CA, AE, MX, SG, AU, NZ, GB, CH, BH.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 26, 2025
Recall initiation date
Feb 6, 2025
Center classification date
Mar 17, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.