epoc BGEM Test Card 25 Test Cards/box. In vitro diagnostic device for the quantitative testing of samples of heparini... — FDA record Z-1372-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jun 1, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
epoc BGEM Test Card 25 Test Cards/box. In vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood Product Siemens SMN#/Model #: 10736515
Reason
Discrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support.
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics Inc
Quantity
263 Boxes
Agency-reported status
Ongoing
Source record identifier
Z-1372-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00809708121860 Lot Number: 11-22005-50 Exp. Date: 22-June-2022

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

AL CA CO FL IA ID KS NM OK TX

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jul 20, 2022
Recall initiation date
Jun 1, 2022
Center classification date
Jul 12, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.