MINI STICK MAX 5F X 15 CM STIFF .018 NI/TU NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire in... — FDA record Z-1347-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 30, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MINI STICK MAX 5F X 15 CM STIFF .018 NI/TU NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-776
Reason
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
Classification
Class II
Recalling firm
Angiodynamics, Inc.
Quantity
40 units
Agency-reported status
Ongoing
Source record identifier
Z-1347-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 15051684023191 UPN: H965457761 Lot Number: 5787840 5803520 5805069

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.